esi SR1010 Resistance Transfer Standard
Specifications
Transfer Accuracy: ±(1 ppm + 0.1 μΩ) at parallel value for 100:1 transfer ±(1 ppm + 1 μΩ) at series-parallel value for 10:1 transfer
Initial Accuracy ±20 ppm of nominal value matched within 10 ppm
Long-Term Accuracy ±50 ppm of nominal value
Calibration Accuracy ±10 ppm
Calibration Conditions 23 ºC, low power, four-terminal measurement
Temperature Coefficient ±5 ppm/ºC matched within 3 ppm/ºC for 100 Ω and higher, ±1 ppm/ºC for 10 Ω ±15 ppm/ºC matched within 5ppm/ºC for 1 Ω
Power Coefficient ±0.1 ppm/mW per resistor for 100 Ω and higher ±0.02 ppm/mW per resistor for 10 Ω ±0.3 ppm/mW per resistor for 1 Ω
Maximum Power Rating 1Ω/step or 5 W distributed over 10 resistors
Breakdown Voltage 1500 V peak to case
Leakage Resistance Greater than 1012 Ω from terminal to case.