Temperature measurement where recorded data needs to be accessed without a PC
The LogTag® TRID30-7 temperature recorder features a display together with data logging function storing up to 7770 temperature readings and a separate rolling 30 day statistical max/min reading and duration memory which can be reviewed on the display.
The visual display of current temperature and previous alarms is an important feature in "static" applications such as cool rooms and refrigerators.
Suitable for applications requiring a visual display of current temperature and history such as coolrooms, refrigerators etc but also can be used in transit applications where a direct readout of the statisitical data is required without use of a computer or readout device.
The display arrangement is designed to provide 'at a glance' if temperature violations have occurred for both the current reading and up to the last 30days.
Details of any excursions can be checked directly by inspecting the statistics history on the recorder's display or in more detail by downloading the logged data. If a reading outside the pre-set "Alarm" limits is recorded at any time, a "day alarm indicator" appears on the display.
Logged data can be downloaded via a standard LogTag® Interface to the companion software LogTag® Analyzer which provides facilities for displaying data in chart, table or statistical formats and allows electronic archiving, export or transmission of the data in support of sophisticated data management systems.
The recorder incorporates an external temperature sensor uniquely arranged in a recess in the case.
This arrangement protects the sensor from physical damage while retaining the faster response time normally associated with remote sensors.
Please note: An interface is required for configuration and reporting
The normal recording display shows the current temperature reading, the time and the alarm trigger day summary. A visual alarm trigger is displayed if one or more of the configured alarm trigger conditions have been met. An alarm trigger condition can be a single reading above the upper or below the lower threshold values, a set of consecutive violating readings or an accumulation of violation readings encountered.
These display examples show alarm triggers recorded 1,8, 20, 21 and 26 days ago. Review each day's statistic by pressing the REVIEW button.
Statistics include the maximum and minimum reading for each day, durations above or below preset thresholds and alarm stati.
The recorder can be configured to suspend processing of readings for alarms and max/min statistics for a period of time after button press activity.
This allows the user to review the current statistics or clear an alarm without causing a false alarm or statistic while handling the recorder.
Part Order Code | TRID30-7R : (replaceable battery CR2032) TRID30-7F : (fixed battery) |
Rated Operating Temperature range | -30 °C ~ +60 °C |
Rated Temperature reading accuracy | Better than ±0.5 °C for measurements from -20 °C~+40 °C -typically ±0.3 °C Better than ±0.8 °C for other measurements - typically ±0.5 °C |
Rated temperature resolution on LCD | Data logging memory : 7770 logs - eg 53 days @ 10min logging, 80 days @ 15 min logging Day summary statistics memory (for display on LCD): up for 30 days of Max/Min and duration values |
Sampling Interval | configurable from 30seconds to hourly |
Logging Modes | Supports 'continuous' wrap around or specific recording period |
Real time clock accuracy | Quartz crystal locked real time clock. Rated accuracy ±25 ppm @ 25 °C (equiv to 2.5 seconds/day) Rated temperature coefficient is -0.034±0.006 ppm/°C (i.e typically +/- 0.00294 seconds/day/°C) |
Download time | typically less than 5 seconds for full memory |
Power Source | 3 V Lithium-Manganese Dioxide extended temperature chemistry |
Battery Life | TRID30-7R: typically 1 year of operation TRID30-7F: typically 2+ years of operation |
Size | 93.0 mm (H )x 54.5 mm (W) x 8.6mm (T) |
Weight | TRID30-7R: 39grams TRID30-7F: 43grams |
Case Material | Polycarbonate |
EMC Compliance | Complies with EC EMC directives (EN50081-1:1992 & EN 61000-6-1:2001) Complies with FCC Part 15 Subparts A and B. |
Environmental | IP65 (when vertically mounted or hung) |
FDA 21CFR Part 11 Compliance | Designed to comply with FDA 21CFR Part 11 (Digital signatures) |
Reader interface | Standard LogTag® Interface cradle |
Manufactured to meet the European RoHS Directive |
We cooperate with partners and resellers in other European countries