Single-Use USB Temperature Recorder with Display
The LogTag® USRID-16W range measures and stores 40 days of real time temperature readings over a measurement range of -30 °C to +60 °C, and is pre-qualified for World Health Organisation (WHO) PQS E006 Standards.
Data logger with integral USB connector. The LogTag® WHO compliant USRID-16WC and USRID-16WA/B satisfy the growing need for a cost effective and reliable single-use temperature recorder with built-in USB and display. They are ideally suited for vaccine transportation, pharmaceutical transportation as well as in laboratories and medical centres & healthcare.
The display shows alarm status of up to 3 alarms set at WHO requirements. It plugs directly into the computer USB for easy data download.
Upon download, it automatically generates a fully detailed PDF report with the option to download the data using the freely available LogTag® Analyzer software for a more detailed analysis.
WHO Type 1 alarm configuration for freeze sensitive vaccines:
DTP, DT
TT, Td
HepB, IPV, HPV
Liquid Hib and combination vaccines
Pneumo (other than Prevenar)
WHO Type 2 alarm configuration suitable for the international shipment of:
BCG, lyophilized Hib, OPV
Measles, MR, MMR, Rabies,
Meningitis (polysaccharide)
Yellow fever vaccines
Rotavirus (other than rotateq)
Product Models | USRID-16WC USRID-16WA/B |
Sensor Measurement Range | -30 °C bis +60 °C |
Operating Temperature Range | -30 °C bis +60 °C |
Storage Temperature Range | 0 °C to +40 °C with device de-activated |
Rated Temperature Reading Accuracy | Better than ±0.5° C for -5 °C to +30 °C Better than ±0.8 °C or better for -30 °C to +60 °C Actual performance is typically much better than the rated values. Accuracy figures can be improved by recalibration. |
Rated Temperature Reading Resolution | Better than 0.1 °C between -20 °C to +55 °C LogTag Analyzer® currently displays to one decimal place of °C. The native resolution is what is stored in the LogTag®. |
Sensor Reaction Time | Typically less than 7 minutes (T90) in moving air (1 m/s) |
Recording Capacity | 11,520 temperature readings (40 days at 5 minute logging) |
Sampling Interval | Factory set at 5 minutes (non-changeable) |
Logging Start Options | Push button start |
Logging Stop Options | Push Button Stop. Stop after the logger has completed 11,520 recordings (40 days at 5 minute logging). |
Alarms | 3 alarms set at WHO requirements |
Download Time | Typically with full memory in less than 20 seconds from time of insertion to availability of PDF report. Typically less than 10 seconds from time of insertion to availability of LTD file in LogTag® Analyzer 3. |
Environmental | Ingress rating: IP65 with USB cap fitted. Shock and Vibration: Unit shall pass a relevant level of shock and vibration based on the requirements of the data logger specification E006/ TR07.3. Environmental resistance: Labels and plastics shall be resistant to a range of chemicals and UV, including rubbing alcohol and petrol fuel. |
Power Source | CR2032 3V LiMnO2 Battery (Non-Replaceable). |
Battery Life | Minimum storage life of 18 months before 'start'. Monitoring period: 40 days. Minimum accessibility (display) period of 6 months after 'stopped'. |
Real Time Clock | Built-in real time clock. Rated accuracy ±25 ppm @ 25 °C (equivalent to 2.5 seconds/day). Rated temperature coefficient is -0.034 ±0.006 ppm/°C (I.e typically +/- 0.00294 seconds/day/°C) |
Connection Interface | USB 2.0, A-type plug. |
Software | PDF Reader, LogTag® Analyzer 3 or higher |
Size | 93 mm(H) x 54.5 mm(W) x 8.6 mm(T) including protective USB cap |
Weight | 31g |
Case Material | Polycarbonate |
We cooperate with partners and resellers in other European countries
View partners